Surtron® Flash 120

The Surtron Flash 120 is an electrosurgical generator which is suitable for precise monopolar and bipolar surgery and micro-surgery without tissue alterations.

Surtron Flash 120 can supply RF operative powers either for programmable time as short as hundredths of a second.

Through its performances, Surtron Flash 120 allows pure CUT,  cut-coagulation BLEND, incision with reduced production of eschar ENHANCED, superficial coagulation FORCED COAG, deep coagulation in absence of necrosis SOFT COAG, BIPOLAR coagulation and ABLATION.

The digital reading of the delivered power and the overseeing through microcontroller of the operational functions, assure the absolute reliability of the conditions of job.

Surtron Flash 120 allows a highly professional surgery thanks to the user-friendly and safety solutions normally used.

The connection of the neutral electrode is constantly monitored. Safety control of patient/plate contact using a split neutral electrode. The possibility to control by the handle the output functions as well as the delivery of output power allows implementing the surgical operation without turning away the surgeon attention from the surgical field.

 

Please be aware that due to our environmental and sustainability commitments, this product is supplied without a power cable.

You must purchase the following in order to use your Surtron® Flash 120: Hospital Grade Power Cable

What's in the box

Surtron Flash 120 diathermy unit

Reusable handle with switches

Assorted electrodes (10 pcs)

Disposable Cable for neutral electrode

Steel neutral electrode

Foot switch

Needles for Microsurgery (5pcs)

Reusable Handle for Microsurgical Needle

Disposable Neutral Electrode (x2) 

Disposable Split Neutral Electrode (x2)

Blade Electrode 7cm (x3)

Needle Electrode 7cm (x3)

Ball Electrode 7cm [⌀4mm] (x3)

Dimensions

Length: 319mm

Width: 370mm

Height: 144mm

Weight: 6kg

Output Power

Max output power CUT 120 W-25 Ω

Max output power BLEND 90 W-20 Ω

Max output power ENHANCED 90 W-250 Ω

Max output power FORCED COAG 80 W-150 Ω

Max output power SOFT COAG 60 W-100 Ω

Max output power BIPOLAR COAG 60 W-50 Ω

Working frequency 600 KHz

RF power pulse emission time 10 ms – 30 sec

RF power pulse interval 1 – 9

Patient circuit F

Selectable input voltage 115 – 230 Vac

Mains frequency 50 – 60 Hz

Electrical input power 300 VA

Safety

EN60601-1
EN60601-1-2
EN60601-2-2
Electrical classification: CF
MDD 93/42 / EC: II b

Download Brochure

Surtron Flash 120 Brochure